CalCheck Negative Control Failing

image_pdfimage_print

Introduction

If when reading an unactivated ATP test, an activated ATP test that hasn’t swabbed a surface or running the negative calibration check with the CalCheck results > 4 RLUs there could be some light leakage into the chamber or some phosphorescing material in the chamber around the sensor. Some good examples are:

1. EnSURE Touch lid is not closing completely and or the lid is damaged.

2. The read chamber plug at the bottom of the EnSURE Touch is not screwed all the way into the base of the unit.

3. Cracks in the case. While the EnSURE Touch has a light-tight inner chamber it is possible cracks of broken parts of the case could allow light into the chamber.
4. Phosphorescing or light-emitting material has gotten into the chamber by the sensor and dried to the surface. This can include:
a. Chemistry from the ATP devices if the device was not used correctly and chemistry from the device was able to get on reading chamber walls. Example: placing the EnSURE Touch on a flat surface while the device is still in the chamber.
b. If the user added their own white labels to the CalCheck control calibration device. (White labels will generate RLU signal)
5. To detect a light entering the chamber have everything ready to run the test and shut the lights off for complete darkness to verify that there is no light entering the chamber.

6. If your unit is still failing please see our How to Clean your instrument related article below and then repeat the test.

If you still need help click here to submit a ticket for technical support.

Tags:CalibrationEnsure Touch
Related Articles
  • How to Register Your EnSURE Touch
  • Add a Sync Server Address
  • Calibration Checks on the EnSURE Touch
  • Settings to Maximize Your EnSURE™ Touch Battery Life Between Charging Events.
  • Languages available on EnSURE Touch luminometer.
  • Accessing Captive Portals on EnSURE Touch
EnSURE™ Touch
  • CalCheck Negative Control Failing
  • How to Register Your EnSURE Touch
  • Add a Sync Server Address
  • Calibration Checks on the EnSURE Touch
  • Settings to Maximize Your EnSURE™ Touch Battery Life Between Charging Events.
  • Languages available on EnSURE Touch luminometer.
View All 32  
Popular Articles
  • Connecting to Wi-Fi
  • Importing Programming Information Into SureTrend Cloud
  • Getting Started with EnSURE Touch
  • Calibration Checks on the EnSURE Touch
  • USB Tethering Configuration
KB Categories
  • ATP
    • Microbial Detection
  • CalCheck
  • Connectivity
  • EnSURE and SystemSURE Plus
  • EnSURE™ Touch
  • Getting Started
  • Hardware
  • Hygiena Academy
  • Reagents, Tests and Devices
  • SureTrend 4
  • SureTrend Cloud
  • SureTrend CloudSync
About Hygiena Help

We stand behind the quality of our products and are committed to our customers. That is why we’re proud to offer world-class support.

Recommended
  • Performing a Quick Test
Most Popular Articles
  • Managing the Security Feature for Ensure Touch™ from the Cloud
  • CalCheck Negative Control Failing
  • Blurred or Distorted EnSURE Touch Screens
  • Excluding Plans and Locations on EnSURE Touch
Latest Articles
  • Moving the Database
  • Server Installation
  • Connecting to Wi-Fi
  • Performing a Quick Test
  • © 2021 Hygiena, LLC.. All Rights Reserved.