• FAQ
  • Manuals
  • Knowledge Base
  • Videos
  • Downloads
  • FAQ
  • Manuals
  • Knowledge Base
  • Videos
  • Downloads
What's Popular:quick test, wifi, syncing

EnSURE™ Touch Only Reads 0 RLU

image_pdfimage_print

Introduction

       If EnSURE™ Touch only reads 0 RLUs with a CalCheck, Calibration Control Kit, or when you intentionally grossly contaminate a Hygiena™ ATP test device (ie. swab your hand) it is likely the shutter has become dislodged or stuck. This can happen if the instrument is dropped without the protective case on or if liquids are trapped in the read chamber, leaked into the shutter compartment and caused the shutter to become stuck. Most likely you need to send the instrument in for repair. However, we would like you to verify the shutter is stuck before sending in. Follow the instructions below to verify the shutter is stuck. 

 

How To Troubleshoot a Stuck Shutter on EnSURE™ Touch

  • Go to Quick Test     
  • Place a Calibration Device (CalCheck or Calibration Control Kit) or Hygiena ATP device (activated or unactivated) into EnSURE Touch and shut lid. 
  • Run Test 
  • During the 10 second count down around the 4-3 second mark and 2-1 second mark you should hear the shutter open and close. If you do not hear any mechanical click, it means the shutter is stuck. 

 

If you still need help click here to submit a ticket for technical support.

 

 

 

Tags:CalibrationEnsure TouchRLU
Related Articles
  • How to Register Your EnSURE Touch
  • Add a Sync Server Address
  • Calibration Checks on the EnSURE Touch
  • Settings to Maximize Your EnSURE™ Touch Battery Life Between Charging Events.
  • Languages available on EnSURE Touch luminometer.
  • Accessing Captive Portals on EnSURE Touch
EnSURE™ Touch
  • EnSURE™ Touch Only Reads 0 RLU
  • How to Register Your EnSURE Touch
  • Add a Sync Server Address
  • Calibration Checks on the EnSURE Touch
  • Settings to Maximize Your EnSURE™ Touch Battery Life Between Charging Events.
  • Languages available on EnSURE Touch luminometer.
View All 32  
Popular Articles
  • Connecting to Wi-Fi
  • Importing Programming Information Into SureTrend Cloud
  • Getting Started with EnSURE Touch
  • Calibration Checks on the EnSURE Touch
  • USB Tethering Configuration
KB Categories
  • ATP
    • Microbial Detection
  • CalCheck
  • Connectivity
  • EnSURE and SystemSURE Plus
  • EnSURE™ Touch
  • Getting Started
  • Hardware
  • Hygiena Academy
  • Reagents, Tests and Devices
  • SureTrend 4
  • SureTrend Cloud
  • SureTrend CloudSync
About Hygiena Help

We stand behind the quality of our products and are committed to our customers. That is why we’re proud to offer world-class support.

Recommended
  • Performing a Quick Test
Most Popular Articles
  • Managing the Security Feature for Ensure Touch™ from the Cloud
  • CalCheck Negative Control Failing
  • Blurred or Distorted EnSURE Touch Screens
  • Excluding Plans and Locations on EnSURE Touch
Latest Articles
  • Moving the Database
  • Server Installation
  • Connecting to Wi-Fi
  • Performing a Quick Test
  • © 2021 Hygiena, LLC.. All Rights Reserved.