EnSURE™ Touch Only Reads 0 RLU

image_pdfimage_print

Introduction

       If EnSURE™ Touch only reads 0 RLUs with a CalCheck, Calibration Control Kit, or when you intentionally grossly contaminate a Hygiena™ ATP test device (ie. swab your hand) it is likely the shutter has become dislodged or stuck. This can happen if the instrument is dropped without the protective case on or if liquids are trapped in the read chamber, leaked into the shutter compartment and caused the shutter to become stuck. Most likely you need to send the instrument in for repair. However, we would like you to verify the shutter is stuck before sending in. Follow the instructions below to verify the shutter is stuck. 

 

How To Troubleshoot a Stuck Shutter on EnSURE™ Touch

  • Go to Quick Test     
  • Place a Calibration Device (CalCheck or Calibration Control Kit) or Hygiena ATP device (activated or unactivated) into EnSURE Touch and shut lid. 
  • Run Test 
  • During the 10 second count down around the 4-3 second mark and 2-1 second mark you should hear the shutter open and close. If you do not hear any mechanical click, it means the shutter is stuck. 

 

If you still need help click here to submit a ticket for technical support.

 

 

 

Tags:CalibrationEnsure TouchRLU
Related Articles
  • Managing Plans & Locations In SureTrend Cloud
  • Adding a New/Replacement EnSURE Touch to a Site
  • How to set up a mobile Hotspot
  • Product Selection instructions for EnSURE Touch
  • How to Factory Reset the EnSURE Touch
  • Setting specific plans and locations on Ensure Touch instruments registered to the same site.
EnSURE™ Touch
  • EnSURE™ Touch Only Reads 0 RLU
  • Managing Plans & Locations In SureTrend Cloud
  • Adding a New/Replacement EnSURE Touch to a Site
  • How to set up a mobile Hotspot
  • Product Selection instructions for EnSURE Touch
  • How to Factory Reset the EnSURE Touch
View All 26  
Popular Articles
  • Connecting to Wi-Fi
  • USB Tethering Configuration
  • Importing SureTrend 4 data into SureTrend Cloud
  • Getting Started with Cloud Sync
  • Protected: Battery Not Charging
KB Categories
  • ATP
  • CalCheck
  • Connectivity
  • EnSURE and SystemSURE Plus
  • EnSURE™ Touch
  • Hardware
  • Reagents, Tests and Devices
  • SureTrend 4
  • SureTrend Cloud
  • SureTrend CloudSync
About Hygiena Help

We stand behind the quality of our products and are committed to our customers. That is why we’re proud to offer world-class support.

Recommended
  • Performing a Quick Test
Most Popular Articles
  • Effects of Environmental Light and Temperature on Hygiena ATP test devices
  • Deactivating Sites in SureTrend Cloud
  • Domain Locks for SureTrend Cloud
  • Adding Zone detail to Locations on the SureTrend Cloud and Ensure Touch
Latest Articles
  • Moving the Database
  • Server Installation
  • Connecting to Wi-Fi
  • Performing a Quick Test
  • © 2019 Hygiena, LLC.. All Rights Reserved.