Optimal Power Management Settings for your EnSURE™ Touch

Optimal Display and Power Settings

  1. From the Home Screen, tap Settings.
  2. Tap General.
  3. Tap Display & Power Off.
  4. Adjust the Display & Power Off settings as listed below.
    • Set Display Brightness to 40% or less.​
    • Set Display Timeout to 1-5 minutes.​
    • Set Power Off After Display Timeout to Off.
    • Set Power Off Daily at to Off.

Power Saving Recommendations

  • Letting the screen sleep between tests or putting the screen to sleep by pressing the power button once will extend the battery charge.​
  • Turn the Wi-Fi off while the device is in use can help to increase the battery life on your luminometer. You only need the Wi-Fi on when syncing, updating the software, or getting remote technical support.
    • Go to Settings > Wi-Fi and tap the toggle to turn Wi-Fi off.
  • Make certain that the EnSURE Touch is properly powered off after use by holding the power for about 1-2 seconds then selecting Power Off from the menu that appears.
  • Enabling the ‘After Display Timeout’ and/or the ‘Daily At’ will result in your luminometer automating the shutdown process thus requiring a power reset sequence to reinitiate the operating system. If pressing the power button does not initiate the operating system, please try a 12 second hold on the power button.

If you are having trouble with the battery life on your EnSURE Touch, please contact Hygiena Technical Support for assistance.