• FAQ
  • Manuals
  • Knowledge Base
  • Videos
  • Downloads
  • Contact Us
  • FAQ
  • Manuals
  • Knowledge Base
  • Videos
  • Downloads
  • Contact Us
What's Popular:quick test, wifi, syncing

Blurred or Distorted EnSURE Touch Screens

image_pdfimage_print

Introduction

The EnSURE Touch has an electrical component that handles the communication between the system and the screen.  This component is sensitive to electromagnetic frequencies, like radio waves and wireless communication signals.  The performance of the component can also be impacted when the internal plastic components that mount the electronics in the instrument become damaged.

When the instrument displays a blurred, pixelated screen, this likely due to RF interference.  The electrical components in the EnSURE Touch are susceptible to radio frequencies ranging from 100MHz to 500MHz.  When RF waves are introduced to the instruments in proximity (i.e. radio or walkie talkie is activated near the instrument), the signals interfere with the communication between the hardware and the screen becomes blurred.  If this problem is occurring on your instrument, see the solutions section for instructions on how to correct the issue.

When the instrument displays a blue or green distorted display, this is likely due to damage to the instrument.

Solutions

  1. Press the power button to turn off the display.
  2. Remove any sources of RF interference like walkie talkies or other radios within proximity of the instrument.
  3. Press the power button to turn on the display.
  4. If the screen remains blurry there may be other sources of RF interference. Try moving to another location where there are no RF signals and repeat this process.
  5. If you are unable to correct the blurred screen using the steps listed here, contact Hygiena Technical support by clicking here.

 

 

Tags:blurredblurringdisplaydistortdistortedpixelatedscreen
Related Articles
  • Managing Plans & Locations In SureTrend Cloud
  • Adding a New/Replacement EnSURE Touch to a Site
  • How to set up a mobile Hotspot
  • Product Selection instructions for EnSURE Touch
  • How to Factory Reset the EnSURE Touch
  • Setting specific plans and locations on Ensure Touch instruments registered to the same site.
Hardware
  • Blurred or Distorted EnSURE Touch Screens
  • Quick Boot vs Normal Boot
  • Protected: Battery Not Charging
  • Battery Replacement
  • Battery Basics
Popular Articles
  • Connecting to Wi-Fi
  • USB Tethering Configuration
  • Importing SureTrend 4 data into SureTrend Cloud
  • Getting Started with Cloud Sync
  • Protected: Battery Not Charging
KB Categories
  • ATP
  • CalCheck
  • Connectivity
  • EnSURE and SystemSURE Plus
  • EnSURE™ Touch
  • Hardware
  • Reagents, Tests and Devices
  • SureTrend 4
  • SureTrend Cloud
  • SureTrend CloudSync
About Hygiena Help

We stand behind the quality of our products and are committed to our customers. That is why we’re proud to offer world-class support.

Recommended
  • Performing a Quick Test
Most Popular Articles
  • Effects of Environmental Light and Temperature on Hygiena ATP test devices
  • Adding Zone detail to Locations on the SureTrend Cloud and Ensure Touch
  • Deactivating Sites in SureTrend Cloud
  • Adding a New/Replacement EnSURE Touch to a Site
Latest Articles
  • Moving the Database
  • Server Installation
  • Connecting to Wi-Fi
  • Performing a Quick Test
  • © 2019 Hygiena, LLC.. All Rights Reserved.